Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients
Author:
Contact Author:

Xin Tang. Tianjin Eye Hospital, Tianjin Key Laboratory of Ophthalmology and Vision Science, Clinical College of Ophthalmology, No. 4 Gansu Road, Heping District, Tianjin 300020, China. professortangxin@163.com

Affiliation:

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To determine the repeatability of Ophtha Top topography and assess the consistency with intraocular lens (IOL)-Master and LenstarLS900 (Lenstar) in measuring corneal parameters among cataract patients. METHODS: Totally 125 eyes were enrolled. Corneas were successively measured with Ophtha Top, IOL-Master and Lenstar at least three times. The flattest meridian power (Kf), the steepest meridian power (Ks), mean power (Km), J0 and J45 were recorded. Intra-class correlation coefficients (ICCs), the coefficient of variance (COV), within subject standard deviation (Sw), and test-retest repeatability (2.77Sw) were adopted to determine the repeatability. The 95% limit of agreement (95%LOA) and Bland-Altman plots were used to assess comparability. RESULTS: Repeatability of Ophtha Top topography for measuring corneal parameters showed the ICCs were all above 0.93, 2.77Sw was lower than 0.31, and the COV of the Kf and Ks was lower than 0.25. The keratometric readings with Ophtha Top topography were flatter than with the IOL-Master and Lenstar devices, while the Pearson correlation coefficients were over 0.97. The J0 and J45 with Ophtha Top topography were smaller compared with Lenstar and IOL-Master, while was comparable between Lenstar and IOL-Master. CONCLUSION: Ophtha Top topography shows excellent repeatability for measuring corneal parameters. However, differences between the Ophtha TOP topography and Lenstar, IOL-Master both in cornea curvature and the astigmatism should be noted clinically.

    Reference
    Related
    Cited by
Get Citation

Sha-Sha Yu, Hui Song, Xin Tang. Repeatability of Ophtha Top topography and comparison with IOL-Master and LenstarLS900 in cataract patients. Int J Ophthalmol, 2017,10(11):1703-1709

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:October 23,2016
  • Revised:December 16,2016
  • Adopted:
  • Online: November 07,2017
  • Published: