Validation and reproducibility of the Heidelberg Edge Perimeter in the detection of glaucomatous visual field defects
Author:
Contact Author:

Jonathan S. Myers. Wills Eye Hospital, 840 Walnut St, Philadelphia PA 19107, USA. jmyers@willseye.org

Affiliation:

Clc Number:

Fund Project:

Supported by Wills Eye Innovation Grant #WEF15064. Heidelberg Engineering Inc. (California, USA) provided the Heidelberg Edge Perimeter used in this study. The sponsor had no role in the design or conduct of this research.

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    AIM: To validate the ability of the Heidelberg Edge Perimeter (HEP) in detecting glaucomatous visual field (VF) defects compared to the Octopus Visual Field (OVF) Analyzer and to determine the test-retest repeatability of both modalities. METHODS: This prospective, cross-sectional study was conducted at Wills Eye Hospital, Philadelphia, PA. Glaucoma subjects and unaffected controls underwent VF testing using HEP standard automated perimetry (SAP) III 30-2 Adaptive Staircase Thresholding Algorithm (ASTA) FAST protocol and OVF G-TOP white-on-white strategy. Testing order was randomized. Pearson’s correlation coefficient was calculated for both mean deviation (MD) and pattern standard deviation/square root of loss of variance (PSD/sLV). Receiver operating characteristic (ROC) curves were used to determine the diagnostic abilities of both modalities. Glaucoma subjects returned for repeat testing and intraclass correlation coefficients (ICCs) were calculated to assess test-retest repeatability. RESULTS: Seventy-nine subjects with glaucoma and 36 unaffected controls were enrolled. HEP and OVF were significantly correlated for both MD (r=-0.84; P<0.01) and PSD/sLV (r=0.79; P<0.01). Areas under the ROC curves (AUCs) were also comparable between HEP and OVF for both MD (0.74 and 0.79, respectively; P=0.26) and PSD/sLV (0.74 and 0.82; P=0.08). ICC was high for both HEP (0.96 for MD, 0.95 for PSD/sLV) and OVF (0.82 and 0.88, respectively). Mean test duration (min) was significantly shorter for OVF (2.63) compared to HEP (5.15; P<0.01). CONCLUSION: HEP and OVF show strong correlation for both MD and PSD/sLV, and have similar validity for detecting glaucoma. The test-retest repeatability is high for both HEP and OVF, however the average testing duration of HEP is significantly longer than that of OVF.

    Reference
    Related
    Cited by
Get Citation

Cui QN, Gogte P, Lam JM, Siraj S, Hark LA, Myers JS, Katz LJ, Waisbourd M. Validation and reproducibility of the Heidelberg Edge Perimeter in the detection of glaucomatous visual field defects. Int J Ophthalmol 2019;12(4):577-581

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:September 18,2018
  • Revised:January 29,2019
  • Adopted:
  • Online: March 06,2019
  • Published: